INFINITY SWEDEHEART: Non inferiority after twelve months, significant benefit from month six
Patients in the largest randomized controlled trial of the DynamX Coronary Bioadaptor.[1]
For the primary endpoint, target lesion failure, within the non inferiority margin of 4.2 percentage points.[3][4]
DynamX Coronary Bioadaptor compared with the comparator stent in the landmark analysis.[2]
With 2,400 patients, INFINITY SWEDEHEART is the largest randomized controlled trial to evaluate the DynamX Coronary Bioadaptor against a contemporary drug eluting stent.[1]
More than 76 percent of enrolled patients presented with acute coronary syndrome.[1] The trial was embedded in the Swedish SWEDEHEART registry, which captures treatment data from routine clinical care.[1] However, patients with left main disease, chronic total occlusion, bypass graft as the target vessel, or in stent restenosis were excluded from the trial.[6]
For the primary endpoint, target lesion failure, the difference after twelve months was 0.41 percentage points compared with the comparator stent.[1][2] This difference fell within the prespecified non inferiority margin of 4.2 percentage points.[3][4] The p value for non inferiority was below 0.001.[3] Target lesion failure is a composite endpoint consisting of cardiovascular death, target vessel myocardial infarction, and clinically indicated target lesion revascularization.[1]
In a separately prespecified landmark analysis between month six and month twelve, the DynamX Coronary Bioadaptor showed a TLF rate of 0.2 percent compared with 1.3 percent for the comparator stent, with a p value of 0.003.[2] Between month six and year two, the TLF rate was 1.47 percent compared with 2.81 percent, corresponding to a hazard ratio of 0.52 and a relative reduction of approximately 48 percent.[7] A significant benefit for the bioadaptor was also seen in the high risk subgroup of patients with acute coronary syndrome during this period.[5]
The difference between the two groups becomes visible from month six onward.[1][7] During this same period, the polymer coating of the DynamX Coronary Bioadaptor begins to resorb, and the unlocking mechanism of the implant is activated.[1]
An implant that unlocks after six months
Share of patients with target lesion failure (TLF)
Non inferiority met, difference 0.41 percentage points, p value below 0.001.[1][3] A difference in the test for superiority was not significant.[6]
Share of patients with target lesion failure (TLF)
p value 0.003.[2]
Share of patients with target lesion failure (TLF)
Hazard ratio 0.52: In this period, the risk was calculated at a little over half.[7]
The DynamX Coronary Bioadaptor is a coronary implant positioned between a conventional drug eluting stent and a fully bioresorbable scaffold.[9] It consists of three helical strands made of cobalt chromium that are connected by a biodegradable polymer.[9] After about six months, the polymer is resorbed, the strands unlock, and the vessel can move more freely again.[1][9]
Contact Angiopro
Please enter your name and the address of your clinic when booking.
You will receive the link to the online meeting by email after booking.
What exactly does unlocking mean for the DynamX Coronary Bioadaptor?
The DynamX Coronary Bioadaptor consists of three helically wound metal strands, the helical strands, which are connected by a thin, biodegradable polymer.[9] This polymer is resorbed within about six months, the strands separate from one another, and the vessel can move more freely again.[1][9]
Is the DynamX Coronary Bioadaptor a fully resorbable implant?
No. It is positioned between a conventional drug eluting stent and a fully bioresorbable scaffold, that is, a structure designed to dissolve completely.[9] Only the connecting polymer is resorbed. The cobalt chromium strands remain in the vessel and continue to support it.[9]
Which patients do the results apply to?
INFINITY SWEDEHEART treated 2,399 patients with a mean age of 68 years, of whom 24 percent were women and about 77 percent had acute coronary syndrome.[8] Exclusions included left main lesions, bypass grafts, in stent restenosis, chronic total occlusions, a left ventricular ejection fraction below 30 percent, and impaired kidney function with an eGFR below 30 milliliters per minute.[8] The trial makes no statement about these groups.
How safe was the DynamX Coronary Bioadaptor in the first year?
Cardiac death occurred in 0.6 percent of patients with the DynamX Coronary Bioadaptor and in 0.5 percent with the comparator stent.[8] Stent thrombosis, that is, a vessel blockage caused by a blood clot in the implant, occurred in 0.7 compared with 0.5 percent and was therefore rare in both groups.[8][9]
How long is dual antiplatelet therapy needed after implantation?
Dual antiplatelet therapy means two medications that prevent blood platelets from clumping together. In the trial, it was continued for at least six months in chronic coronary syndrome and at least twelve months in acute coronary syndrome.[8] The duration in an individual case is determined by the treating physician based on the instructions for use and the guidelines.
How robust is the benefit from month six onward?
The benefit comes from a landmark analysis. It counts only events that occur after a defined point in time, here after month six.[2][7] In the test for superiority after twelve months, no significant difference was seen.[6] PCRonline also points out that the event rate in the comparator arm was lower than assumed, which made the non inferiority margin very wide.[6] According to the trial summary, further investigation is needed, including the planned five year follow up.[8]
- INFINITY SWEDEHEART trial page
- ESC Congress 2024: INFINITY SWEDEHEART meets the primary endpoint
- The Lancet: publication of the trial results
- ScienceDirect: trial publication
- Late breaking results, press release of October 27, 2025
- PCRonline: one year results for the primary endpoint
- TCTMD: Unlocking device bests drug eluting stent over time
- American College of Cardiology: INFINITY SWEDEHEART trial summary
- TCTMD: Unlocking scaffold matches drug eluting stent in a wide range of PCI patients, including ACS